Initial Comments:
Based on the findings of an onsite unannounced Medicare recertification survey completed on 8/17/2022, Tyrone Dialysis was found to be in compliance with the requirements of 42 CFR, Part 494.62, Subpart B, Conditions for Coverage of Suppliers of End-Stage Renal Disease (ESRD) Services - Emergency Preparedness.
Plan of Correction:
Initial Comments:
Based on the findings of an onsite unannounced Medicare recertification survey completed 8/17/2022, Tyrone Dialysis was found to have the following standard level deficiencies that were determined to be in substantial compliance with the following requirement(s) of 42 CFR, Part 494, Subparts A, B, C, and D, Conditions for Coverage of Suppliers of End-Stage Renal Disease (ESRD) Services.
Plan of Correction:
494.40(a) STANDARD CARBON ADSORP-MONITOR, TEST FREQUENCY Name - Component - 00 6.2.5 Carbon adsorption: monitoring, testing freq Testing for free chlorine, chloramine, or total chlorine should be performed at the beginning of each treatment day prior to patients initiating treatment and again prior to the beginning of each patient shift. If there are no set patient shifts, testing should be performed approximately every 4 hours.
Results of monitoring of free chlorine, chloramine, or total chlorine should be recorded in a log sheet.
Testing for free chlorine, chloramine, or total chlorine can be accomplished using the N.N-diethyl-p-phenylene-diamine (DPD) based test kits or dip-and-read test strips. On-line monitors can be used to measure chloramine concentrations. Whichever test system is used, it must have sufficient sensitivity and specificity to resolve the maximum levels described in [AAMI] 4.1.1 (Table 1) [which is a maximum level of 0.1 mg/L]. Samples should be drawn when the system has been operating for at least 15 minutes. The analysis should be performed on-site, since chloramine levels will decrease if the sample is not assayed promptly.
Observations:
Based on a review of facility policy and procedure, direct observation (OBS) of staff (EMP) and Staff (EMP) interview the facility staff failed to implement the chlorine testing as per facility policy and procedure for two (2) of two (2) observed and interviewed working (EMP3 and EMP4).
Findings Included:
A review of the agency policies and procedure on 8/11/2022 at approximately 11:00 AM revealed, " ...TITLE: TOTAL CHLORINE TEST USING RPC ULTRA LOW TOTAL CHLORINE TEST STRIP...Procedure 1. Rinse the sample collection cup thoroughly with water to be tested. Fill the sample cup to the line marked 100 ml with the sample. 2. Remove one test strip from the foil package and dip it in the water sample for 60 seconds. While dipping the strip, move it back and forth at a constant gentle rate of approximately two, 1-2 inch wide strokes per second. 3. Remove the strip and shake once, briskly, to remove excess water. 4. Wait 20 seconds for the test strip to develop. While waiting, fold the white plastic handle of the test strip under the test pad area so that it provides a consistent viewing background. As an alternative to folding the strip the reagent pad can be placed directly on the white surface of the color chart when comparing colors. 5. After the 20 second wait period, immediately compare the strip color to the K100-0118 color chart to determine the Total Chlorine level in the sample. NOTE: The strip should yield a uniform color across the entire viewing window. If the window is not a uniform color (e.g., part green and part yellow) repeat the test. A valid result is one in which the viewing window has a uniform color ONLY. 6. Document tests and take any necessary action s per the policy: Daily Water System Total Chlorine Monitoring ...As appropriate, Total Chlorine test results are documented in either the Daily Water Treatment Log-Routine Total Chlorine Testing Log or the Breakthrough Total Chlorine Testing Log. Water used for dialysis contains less than or equal to 0.1 ppm(mg/L) total chlorine (<0.1 ppm). Specific actions are needed if test results are greater than 0.1 ppm (mg/L) (>0.1 ppm)..."
Observation (OBV1) of staff (EMP3) performing Total Chlorine testing using RPC Ultra-Low Total Chlorine Test Strips was conducted on 8/10/2022 at 8:03 AM. EMP3 placed the testing strip into the water sample rapidly moved the strip back and forth for approximately one minute, shook the test strip to remove additional water, then removed the strip from the sample water folded the test strip in half and immediately (approximately within 3 seconds) compared the color with the color chart. The employee did not wait the 20 seconds required by the agency policy.
(EMP1) accompanied the surveyor for (OBV2) per surveyor request. Observation (OBV2) of staff (EMP4) performing Total Chlorine testing using RPC Ultra-Low Total Chlorine Test Strips was conducted on 8/10/2022 at 11:35 AM. EMP4 placed the testing strip into the water sample rapidly moved the strip back and forth for approximately one minute, shook the test strip to remove additional water, then removed the strip from the sample water folded the test strip in half and immediately (approximately within 3 seconds) compared the color with the color chart. The employee did not wait the 20 seconds required by the agency policy.
At 2:20 PM on 8/10/2022 a request for documentation for (EMP3) (EMP4) total chlorine test training within the last year was made from EMP1, EMP2 and EMP6.
During an interview on 8/10/2022 at approximately 2:00 to 2:35 PM both EMP3 and EMP4 confirmed being trained on Chlorine testing but could not verify the exact dates.
The following training documentation was provided for (EMP3): Per training documents, " WATER TREATMENT P&P REVIEW ...Routine Total Chlorine Testing Log, Total Chlorine Breakthrough Testing Log, CM130 Chlorine Test Using ...RPC ultralow Total Chlorine Strip ... " signed facility administrator on 1/19/2022. Also provided was a diploma for completion of " CEC2007: Water Wisdom on 3/3/2022. "
The following training documentation was provided for (EMP4): Per training documents, Total Chlorine Testing and CM130 Operation, if applicable Verbalized reasons for Total Chlorine testing and the consequences of abnormal results. And section Total Chlorine Testing of the document. Signed and dated 1/7/2022. Also provided was a diploma for completion of " CEC4012 Basic Components and Monitoring of Dialysis Quality Water System on 4/8/2022. " An interview was conducted with the administrator on 8/9/2022 at approximately 3:00 PM which confirmed the findings.
Plan of Correction:The Facility Administrator or designee held mandatory in-services for all clinical Teammates starting on 08/17/22. Surveyor observations were reviewed. Education included but was not limited to a review of Policy 02-05-02 "Daily Water System Total Chlorine Monitoring" and Policy 2-05-02H "Total Chlorine Test using RPC Ultra Low Total Chlorine Test Strip" with emphasis on but not limited to: 1) Rinse the sample collection cup thoroughly with the water to be tested. Fill the sample cup to the line marked 100 ml with the sample. 2) Remove one test strip from the foil package and dip it in the water sample for 60 seconds. While dipping the strip, move it back and forth at a constant gentle rate of approximately two, 1-2 inch wide strokes per second. 3) Remove the strip and shake once, briskly, to remove excess water. 4) Wait 20 seconds for the test strip color to develop. While waiting, fold the white plastic handle of the test strip under the test pad area so that it provides a consistent viewing background. 5) After the 20 second wait period, immediately compare the strip color to the K100-0118 color chart to determine the Total Chlorine level in the sample. 6) Document tests results and take any necessary actions per the policy: "Daily Water System Total Chlorine Monitoring". Verification of attendance at in-service will be evidenced by teammates signature on in-service sheet. The Facility Administrator will perform observational audits to verify teammates, including those observed by surveyor, are performing total chlorine monitoring per policy: daily on treatment days for two (2) weeks, then weekly for two (2) weeks, and monthly for two (2) months. Instances of non-compliance will be addressed immediately. Repeat non-compliance will require additional education and possible disciplinary attention. The Facility Administrator will review the audit results with teammates during homeroom meetings, and with the Medical Director during monthly Quality Assessment Performance Improvement meetings known as Facility Health Meetings, with supporting documentation included in the meeting minutes. The Facility Administrator is responsible for compliance with this plan of correction.
494.50(b)(1) STANDARD PERSONNEL HEALTH MONITORING RECORDS Name - Component - 00 4 Records 4.4 Personnel health monitoring records A file must be kept of the results of medical examinations of personnel that are required by OSHA or other regulatory agencies.
Observations:
Based on a review of the agency policy, and staff interview, the agency failed to conduct tuberculosis (TB) screening in accordance with facility policy for one (1) of three (3) PFs reviewed (PF3).
Findings included:
A review of the agency policies and procedure on 8/11/2022 at approximately 11:00 AM revealed, TITLE: TUBERCULOSIS MONITORING AND FOLLOW-UP ...Note: Teammates will adhere to any additional requirements related to initial and annual testing if required by their hospital/institution or local/state health departments. Baseline new hire requirements for all new teammates including volunteers, per diem teammates, non ...agency personnel and teammates will complete the following: 1. TN-Risk Assessment and Symptom Evaluation Questionnaire. 2. Successful completion of Tuberculosis Education for New Teammates course. .. Testing options (any one of the following) ...c. If exemption criteria for TST is not met, the following testing options are available: I. Baseline TST using a twostep Purified Protein Derivative (PPD) Mantoux test (a *second TST repeated one to three weeks after the first, if the initial test is negative). Test results will be recorded on the Teammate Health Monitoring Record (attached) ... "
A review of PF3, date of hire 7/5/2022, was conducted on 8/9/2022 at approximately 2:35 PM. The initial TST was administered on 7/11/2022 the results were read 7/13/2022 with negative results. The agency failed to provide documentation to confirm the second TST was administered within facility policy.
An interview was conducted with the administrator on 8/9/2022 at approximately 3:00 PM which confirmed the findings.
Plan of Correction:The Facility Administrator held mandatory in-services for all clinical teammates starting on 08/07/22. Surveyor observations were reviewed. Education included but was not limited to a review of Policy 4-06-05 "Tuberculosis (TB) Monitoring and Follow Up" with the emphasis on but not limited to: 1) NOTE: Teammates will adhere to requirements related to initial and annual testing if required by their institution or local/state health departments. 2) Baseline new hire requirements for all new teammates including volunteers, per diem teammates, non-agency personnel and teammates will complete the following: 1. TN-Risk Assessment and Symptom Evaluation Questionnaire. 2. Successful completion of Tuberculosis Education for New Teammates course. 3. Testing options ... c. If exemption criteria for TST is not met, the following testing options are available: i. Baseline TST using a two-step Purified Protein Derivative (PPD) Mantoux test (a second TST repeated one to three weeks after the first, if the initial test is negative). Test results will be recorded on the Teammate Health Monitoring Record. Verification of attendance at in-service will be evidenced by teammate's signature on in-service sheet. The Facility Administrator will audit one hundred percent (100%) of new teammate's medical records monthly for two (2) months to verify screening is completed and documentation recorded per policy. Instances of non-compliance will be addressed immediately. The Facility Administrator or designee will review the audit results with the Medical Director during monthly Quality Assessment Performance Improvement meetings known as Facility Health Meetings, with supporting documentation included in the meeting minutes. The Facility Administrator is responsible for compliance with this plan of correction.
494.80(a)(2) STANDARD PA-APPROPRIATENESS OF DIALYSIS RX Name - Component - 00 The patient's comprehensive assessment must include, but is not limited to, the following:
(2) Evaluation of the appropriateness of the dialysis prescription,
Observations:
Based on a review of facility policies and procedures, medical records (MR) and staff (EMP) interviews, the facility failed to ensure the evaluation of the appropriateness of the dialysis prescription including but not limited to: dialysate flow rate (DFR) for one (1) of four (4) in-center hemodialysis medical records (MR) reviewed (MR2).
Findings include:
Review of facility policies conducted on 8/10/2022 at approximately 3:00 PM revealed: "TITLE: PRE-INTRA-POST TREATMENT DATA COLLECTION, MONITORING AND NURSING ASSESSMENT...2. The Nursing assessment will be performed and documented by a licensed nurse; specifically, a Registered Nurse (RN) or if performance of a nursing assessment is permitted by state law, a Licensed Practical Nurse (LPN)/ Licensed Vocational Nurse (LVN). a. The assessment includes the following components...iii Verification of prescription including machine parameters...3. Patient identity, prescription and machine settings are verified by teammate prior to initiation of treatment with the exception of blood flow rate which is verified and documented when the ordered rate is obtained after onset of treatment. The prescription components are confirmed by a licensed nurse within one (1) hour of treatment initiation along with the nursing assessment or as allowable by state law. Prescription components include but are not necessarily limited to: a. Dialyzer make and model b. Treatment time c. Target weight d. UFR (ultrafiltration rare) and Max UFR e. UF Profiling f. Blood flow rate g. Dialysate flow rate..."
MR #2 admit date 12/3/2014, was reviewed on 8/9/2022, at approximately 11:30 AM. A review of the treatment details report dated 8/1/2022 revealed the physician ordered for the DFR at 600 ml/min. DFR was delivered higher than prescribed on 8/1/2022 at 700 ml/min for the entire treatment.
An interview was conducted with the administrator on 8/9/2022 at approximately 3:00 PM which confirmed the findings.
Plan of Correction:The Facility Administrator or designee held mandatory in-services for all clinical teammates starting on 09/12/22. Surveyor observations were reviewed. Education included but was not limited to a review of Policy 1-03-08 "Pre- Intra- Post treatment Data Collection Monitoring and Nursing Assessment" with emphasis on but not limited to: 1) Patient identity, prescription and machine settings are verified by teammates prior to initiation of treatment with the exception of blood flow rate (BFR) which is verified and documented when the ordered rate is obtained after onset of treatment. The prescription components are confirmed by a licensed nurse within 1 hour of treatment initiation along with the nursing assessment. 2) Intra dialytic treatment monitoring and data collection which may be performed by the PCT or licensed nurse includes vital signs and treatment monitoring at least every 30 minutes. At a minimum, obtain and document the following: blood and dialysate flows... 3) If the dialysis prescription is not being met [including dialysis flow rate or change to/inability to obtain prescribed blood flow rate] the reason will be documented and the licensed nurse informed. 4) The licensed nurse notifies the physician (or NPP if applicable) as needed of changes in patient status. 5) All findings, interventions and patient response will be documented in the patient's medical record. Verification of attendance at in-service will be evidenced by teammate's signature on in-service sheet. The Facility Administrator or designee will conduct audits to verify complete and accurate documentation, with notification of and response by the licensed nurse, per policy: on twenty five percent (25%) of the flow sheets daily on treatment days for two (2) weeks then weekly for two (2) weeks. Ongoing compliance will be monitored with the monthly ten percent (10%) medical records audits. Instances of non-compliance will be addressed immediately. The Facility Administrator will review the audit results with teammates during homeroom meetings, and with the Medical Director during monthly Quality Assessment Performance Improvement meetings known as Facility Health Meetings, with supporting documentation included in the meeting minutes. The Facility Administrator is responsible for compliance with this plan of correction.
494.80(a)(2) STANDARD PA-ASSESS B/P, FLUID MANAGEMENT NEEDS Name - Component - 00 The patient's comprehensive assessment must include, but is not limited to, the following:
Blood pressure, and fluid management needs.
Observations:
Based on review of facility policy, medical records (MR) and interview with facility staff, the facility failed to assess and notify appropriate staff for two (2) of four (4) in center hemodialysis medical records reviewed (MR3 and MR4).
Findings Included:
Review of facility policies conducted on 8/10/2022 at approximately 3:00 PM revealed: " TITLE: PRE-INTRA-POSTTREATMENT DATA COLLECTION, MONITORING AND NURSING ASSESSMENT ...2. The Nursing assessment will be performed and documented by a licensed nurse; specifically, a Registered Nurse (RN) or if performance of a nursing assessment is permitted by state law, a Licensed Practical Nurse (LPN)/ Licensed Vocational Nurse (LVN). aa. Vital signs and treatment monitoring i. For non-nocturnal treatments is completed at least every thirty (30) minutes...At a minimum, obtain and document the following: i. Blood Pressure ii Health or pulse rate iii Blood and dialysate flows, arterial and venous pressures iv Fluid removal and/or replacement v. Vascular access visible and line connections intact vi. Patient subjective statements(s) i.e., patient report on overall health, any complaints...11. Abnormal findings or findings outside of any patient specific physician ordered parameters will be reported to the licensed nurse immediately (refer to "Abnormal Findings" section in this policy). The licensed nurse will use his/her clinical judgement based on individual patient needs to determine if any clinical interventions are necessary. 12. The licensed nurse notifies the physician (or NPP if applicable) as needed of changes in patient status...ABNORMAL FINDINGS: Unless other abnormal parameters are established by the facility Governing Body Meeting Minutes, the following are considered abnormal findings and should be reported to the licensed nurse and documented in the patient ' s medical record ...Members of the patient care team should report any changes in patient conditions or concerns of patient well-being immediately to the licensed nurse at any time... Blood pressure: Pre dialysis: *Systolic greater than 180 mm/Hg or less than 90 mm/Hg *Diastolic greater than or equal to 100 mm/Hg Blood Pressure-Intradialytic: *Difference of 20 mm/Hg increase or decrease from patient ' s last Intradialytic treatment BP reading ...Blood Pressure Post Treatment *If the patient can stand: Standing systolic BP greater than 140 mm/Hg or less than 90 mm/Hg Standing diastolic BP greater than 90 mm/HG or less than 50 mm/Hg *If patient is not able to stand, document reason and sitting BP. Sitting BP for patient ' s that cannot stand: Sitting systolic BP greater than 140 mm/Hg or less than 90 mm/Hg Sitting diastolic BP greater than 90 mm/Hg or less than 50 mm/Hg ...Heart or Pulse Rate Pre/Intra/Post: Less than 60 beats per minute or greater than 100 beats per minute and/or an irregular heart beat..."
MR3 review was conducted on 8/9/2022 at approximately 12:00 PM, admission date 8/17/2021, primary diagnosis, end stage renal disease. A review of five (5) treatment records from 7/25/2022 to 8/5/2022.
The following hemodialysis flowsheet dated 7/29/2022 revealed the following: 1:52 111/49 Pulse 104 Documentation by three staff members was completed on the treatment sheet of 7/29/2022.
MR4 review was conducted on 8/9/2022 at approximately 1:00 PM, admission date 5/200/2021, primary diagnosis, end stage renal disease. A review of six (6) treatment records from 7/25/2022 to 8/5/2022.
The following hemodialysis flowsheet dated 7/29/2022 revealed the following: 10:25 BP Pulse 104 10:55 BP 177/101 Documentation by three staff members was completed on the treatment sheet of 7/29/2022.
The following hemodialysis flowsheet dated 7/29/2022 revealed the following: 10:25 Pulse 104 10:55 BP 177/101 Documentation by three staff members was completed on the treatment sheet of 7/29/2022.
The following hemodialysis flowsheet dated 8/5/2022 revealed the following: 10:21 BP 173/105 12:35 BP 179/101 1:21 BP 149/112 1:51 BP 169/104 2:21 BP 165/101 Documentation by three staff members was completed on the treatment sheet of 8/5/2022.
The facility failed to provide documentation to confirm the physician was notified of findings considered abnormal per patients' blood pressure and pulse rate outside the facility parameters.
An interview was conducted with the administrator on 8/9/2022 at approximately 3:00 PM which confirmed the findings.
Plan of Correction:The Facility Administrator or designee held mandatory in-services for all clinical teammates starting on 09/12/22. Surveyor observations were reviewed. Education included but was not limited to a review of Policy 1-03-08 "Pre- Intra- Post treatment Data Collection Monitoring and Nursing Assessment" with emphasis on but not limited to: 1) Patient data will be obtained and documented by the patient care technician or licensed nurse. Data collection includes: Measurement of blood pressure, sitting and standing and intradialytic BP in the sitting/reclined or supine position and pre and post patient weight. 2) Abnormal findings or findings outside of any patient specific physician ordered parameters will be documented and reported to the licensed nurse immediately. The licensed nurse will use his/her clinical judgment based on individual patient needs to determine if any clinical interventions are necessary. 3) The licensed nurse notifies the physician (or NPP if applicable) as needed of changes in patient status. 4) All findings, interventions and patient response will be documented in the patient's medical record. Verification of attendance at in-service will be evidenced by teammates signature on in-service sheet. The Facility Administrator or designee will conduct audits to verify complete and accurate documentation, with notification of and response by the licensed nurse, per policy: on twenty five percent (25%) of the flow sheets daily on treatment days for two (2) weeks then weekly for two (2) weeks. Ongoing compliance will be monitored with the monthly ten percent (10%) medical records audits. Instances of non-compliance will be addressed immediately. The Facility Administrator will review the audit results with teammates during homeroom meetings, and with the Medical Director during monthly Quality Assessment Performance Improvement meetings known as Facility Health Meetings, with supporting documentation included in the meeting minutes. The Facility Administrator is responsible for compliance with this plan of correction.
494.90(a)(1) STANDARD POC-MANAGE VOLUME STATUS Name - Component - 00 The plan of care must address, but not be limited to, the following: (1) Dose of dialysis. The interdisciplinary team must provide the necessary care and services to manage the patient's volume status;
Observations:
Based on reviews of facility policy, medical records (MR) and staff (EMP) interview, the facility failed to provide the care and services necessary to maintain the patient's volume status for one (1) of four (4) in center hemodialysis medical records reviewed (MR1).
Findings Included:
" Unless other abnormal parameters are established by the facility Governing Body and documented in the Governing Body Meeting minutes, the following are considered abnormal findings and should be reported to the licensed nurse and documented in the patient ' s medical record. Within each category, definitions may be adjusted by a patient specific physician order. In addition, the teammate who is observing or collecting information should report to the licensed nurse whenever there is concern for the patient ' s condition or the potential safety of initiating dialysis, even in the absence of specific abnormal findings. Members of the patient care team should report ANY changes in patient conditions or concerns of patient well-being immediately to the licensed nurse at any time. Fluid Status: Pre-treatment: Any weight loss from last post weight, If unable to obtain patient weight, the nurse is to contact physician for orders, Removal goal not to exceed maximum ordered by physician. Post-treatment: If patient is above or below 1 kg from the target weight ... "
MR1, admission date 3/11/2013, was reviewed on 8/9/2022 at approximately 11:15 AM. Per patient summary report section active treatment orders " Prescribed Target Weight 73.5 kg " and ordered maximum ultrafiltration (UF) "Max UF 13 mL/kg/hr" date of order 6/13/2022.
The following treatment records revealed:
7/25/2022 Pre Weight 79.90 kg Scale Post weight 75.10 kg (patient 1.60 kg over ordered target weight 73.5 kg (162.04 lbs) Ordered Maximum Ultra Filtration (UF) 13.0 ml/kg/hr, the actual UFR listed under " POST TREATMENT WEIGHT..Acudal UFR ...25.2 ml/kg/hr. "
7/27/2022 Pre Weight 79.10 kg Scale Post weight 74.50 kg. Ordered Maximum Ultra Filtration (UF) 13.0 ml/kg/hr, the actual UFR listed under " POST TREATMENT WEIGHT. Actual UFR ...21.9 ml/kg/hr. "
7/29/2022 Pre Weight 79.10 kg Scale Post weight 74.20 kg. Ordered Maximum Ultra Filtration (UF) 13.0 ml/kg/hr, the actual UFR listed under " POST TREATMENT WEIGHT. Actual UFR ...22.00 ml/kg/hr. "
8/3/2022 Pre Weight 78.70 Scale Post weight 74.50 kg. Ordered Maximum Ultra Filtration (UF) 13.0 ml/kg/hr, the actual UFR listed under " POST TREATMENT WEIGHT ..Actual UFR ...27.70 ml/kg/hr. "
7/25/2022 Pre Weight 80.10 kg Scale Post weight 75.50 kg (patient 2.0 kg over ordered target weight 73.5 kg (162.04 lbs) Ordered Maximum Ultra Filtration (UF) 13.0 ml/kg/hr, the actual UFR listed under " POST TREATMENT WEIGHT..Acudal UFR ...24.2 ml/kg/hr. "
No documentation was available to confirm the physician notification of target weight greater than 1 (+) (-)-1 kg post treatment, or of (UF) greater than 13.0 mL/kg/hr.
An interview was conducted with the administrator on 8/9/2022 at approximately 3:00 PM which confirmed the findings.
Plan of Correction:The Facility Administrator or designee held mandatory in-services for all clinical teammates starting on 09/12/22. Surveyor observations were reviewed. Education included but was not limited to a review of Policy 1-03-08 "Pre- Intra- Post treatment Data Collection Monitoring and Nursing Assessment" and Policy 1-14-01 "Interdisciplinary Team (IDT) Patient Assessment and Plan of Care" emphasizing but not limited to: 1. Data Collection: 1) Any abnormal findings or findings outside of any patient specific physician ordered parameters discovered during pre-treatment data collection will be documented and immediately reported to the licensed nurse ... If an abnormal finding is reported to the licensed nurse pre-treatment, the nurse will assess the patient prior to the initiation of dialysis. 2) The licensed nurse will use his/her clinical judgment based on individual patient needs to determine if any clinical interventions are necessary. The physician (or non-physician practitioner (NPP) if applicable) will be notified of any concerns that may preclude the initiation of dialysis. 3) The PCT or licensed nurse will obtain and document basic data on each patient post dialysis and compare to pre dialysis findings. 4) If an abnormal finding(s) or concern is identified post treatment, this needs to be reported to the licensed nurse. The licensed nurse will assess the patient prior to discharge. 5) Unless other abnormal parameters are established by the facility Governing Body and documented in the Governing Body Meeting minutes, the following are considered abnormal findings and should be reported to the licensed nurse and documented in the patient's medical record. Fluid Status: Pre-treatment: Removal goal not to exceed maximum ordered by physician. Post treatment: If the patient is above or below one (1) kg from the target weight. 2. IDT: 1) The interdisciplinary team is responsible for providing each patient with an individualized and comprehensive assessment documenting his/her needs... 2) Assessment criteria will include, but not be limited to, evaluation of: Dialysis prescription, blood pressure and fluid management needs. 3) The adequacy of the patient's dialysis prescription will be assessed on an ongoing basis as follows: For hemodialysis patients, at least monthly calculations of delivered Kt/V or an equivalent measure. Verification of attendance is evidenced by teammate's signature on the in-service sheet. The IDT will assess and update the patient care plan for MR1 within thirty days, to address post treatment target weight. The Facility Administrator or designee will conduct flow sheet audits for documentation of post treatment patient weight above or below target weight on twenty five percent (25%) of the flow sheets: daily for two (2) weeks, then weekly for two (2) week. Ongoing compliance will be monitored with monthly ten percent (10%) medical records audits. Instances of non-compliance will be addressed immediately. The Facility Administrator or designee will review audit results with teammates during homeroom meetings, and with the Medical Director during monthly Quality Assessment Performance Improvement meetings known as Facility Health Meetings, with supporting documentation included in the meeting minutes. The Facility Administrator is responsible for compliance with this plan of correction.
|